Performance
- Maximum Heater Plate Temperature 185 deg F (85 deg C)
- Temperature Cut-Out 199 def F (93 def C)
- Maximum Output Humidity 95% relative humidity (control dial set to maximum setting, flow rate 50 L/min, ambient temperature of 68 def F (20 deg C), ambient relative humidity of 40%)
Dimensions (H x W x D)
- Chamber and Docking Station 4.6" x 6.7" x 8.3" (118 mm x 164 mm x 145 mm)
Weight
- Docking Station and Unfilled Water Chamber 670 g (1.5 lb). Water capacity to maximum fill line: 385 mL
Power Supply Input Range
100-240V, 50/60Hz; 110V, 400Hz; 2.5A < 140 VA (110 W) (maximum power consumption).
Instantaneous peak power consumption <340 VA. Actual average power consumption will vary, depending on factors such as temperature setting, set pressure, the use of accessories, height above sea level and ambient temperature.
Maximum heater element power: 85 W.
Housing Construction
- Docking Station Flame retardant engineering thermoplastic, plated cast aluminium. Water chamber: Injection molded plastic, stainless steel.
Environmental Conditions
- Operating Temperature +41 deg F to +104 deg F
- Operating Humidity 10-95% non-condensing
- Storage and Transport Temperature -4 deg F to +140 deg F
- Storage and Transport Humidity 10-95% non-condensing
IEC 60601-1 Classification
Class II (double insulation)
Type CF
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential commercial and light industry environments.
*Consult the medical department of the carrier before using on aircraft.
|